Portable Vision Screener plusoptiX S12C
Plusoptix provides binocular, hand-held vision screening devices, especially for infants, children and uncooperative patients.
Studies show that the best visual outcomes depend on the early detection and treatment of vision disorders. Therefore, paediatric ophthalmologists recommend annual vision screening for all children starting ideally at the age of one year. If there is a family history of eye disorders, earlier screening is advised at the age of 6 to 8 months old.
Conventional visual acuity testing can only be performed in older, cooperative children.
Two healthy eyes are critical to a child’s well-being and development. A visual disorder compromises social and scholastic achievement, can be dangerous in traffic and restrict occupational choices later in life.
The Plusoptix device provides an accurate measurement within seconds. Measurement results are compared to age-specific thresholds, and a “Pass” or “Refer” vision screening result is displayed.
A “Refer” vision screening result will mean the patient should consult an eye care professional for further advice.
This device has a built-in database system, that allows you to populate the kids’ details prior to performing a measurement.
Plusoptix Screening Devices
The Plusoptix S12 range of devices detects refractive errors and eye misalignment, which are the most common causes of amblyopia (lazy eye).
Measurements are non-invasive, requiring no physical contact with the child. A sound target attracts fixation, and no further cooperation is required to perform a measurement.
Features
The measuring principle is based on eccentric photoretinoscopy: infrared light is projected through the pupils onto the retina. Depending on the refractive error, the reflected light forms a specific brightness pattern within the pupil. The spherical refraction is calculated based on this crescent pattern. To determine cylinder and axis, the same measurement is repeated in three meridians. The measurement with infrared light is completely innocuous. Infrared light is also contained in daylight and not visible.
Quality Guaranteed
Documentable
Accurate
Reliable
Studies with plusoptiX Vision Screener
Year | Title | Reference | Results |
---|---|---|---|
2014 | Calibration and Validation of the 2WIN Photoscreener compared to the plusoptiX S12 and the SPOT | JPOS; e-publication | “The plusoptiX outperformed the SPOT and 2WIN as an autorefractor, particularly with respect to astigmatism power and axis.”
2WIN A: 2WIN B: SPOT: plusoptiX S09: plusoptiX S12: |
2013 | A comparison of referral criteria used by the plusoptiX photoscreener | A comparison of referral criteria used by the plusoptiX photoscreener | plusoptiX Criteria 1: Sensitivity 98% Specificity 80%plusoptiX Criteria 2: Sensitivity 81% Specificity 96%plusoptiX Criteria 3: Sensitivity 81% Specificity 92%plusoptiX Criteria 4: Sensitivity 80% Specificity 94%plusoptiX Criteria 5: Sensitivity 98% Specificity 41%plusoptiX Criteria 6: Sensitivity 74% Specificity 86%plusoptiX Criteria 7: Sensitivity 67% Specificity 96% |
2013 | Flip chart visual acuity screening compared to the plusoptiX S09 photoscreener performed by a lay screener | ARVO 2013 Poster # 2517 | Flip chart: Sensitivity 83% Specificity 47%plusoptiX: Sensitivity 92% Specificity 88% |
2012 | Performance of the detection of amblyopia risk factors in children 0 to 5 in central Iowa | AAPOS 2012 Poster # 56 | plusoptiX: Sensitivity 86,8% Specificity 88,0% |
2011 | Validity of plusoptiX S04 photoscreener as a vision screening tool in children with intellectual disability | J AAPOS 2011;15:476-479 | plusoptiX: Sensitivity 95% Specificity 50% |
2011 | Validity of plusoptiX S04 photoscreener as a vision screening tool in children with intellectual disability | AAPOS 2011 Poster # 47 | plusoptiX: Sensitivity 80% Specificity 95% |
2011 | Flip chart visual acuity screening compared to the plusoptiX A09 photoscreener performed by a screener | ESA 2011 Poster # 2001 | Flip chart: Sensitivity 83% Specificity 44%plusoptiX: Sensitivity 94% Specificity 89% |
2010 | Performance of the plusoptiX S04 photoscreener for the detection of amblyopia risk factors in children aged 3 to 5 | J AAPOS 2010;14:147-149 | plusoptiX: Sensitivity 99% Specificity 82% |
2009 | Field testing of the plusoptiX S04 photoscreener | J AAPOS 2009;13:51-57 | plusoptiX: Sensitivity 83% Specificity 95% |
2009 | Comparison between the plusoptiX and MTI Photoscreeners | Arch Ophthalmol.2009;127(12):1591-1595 | MTI: Sensitivity 83,6% Specificity 90,5%plusoptiX: Sensitivity 98,9% Specificity 96,1% |
2008 | Performance of the plusoptiX vision screener for the detection of amblyopia risk factors in children | J AAPOS 2008;12:490-492 | plusoptiX Criteria 1: Sensitivity 98% Specificity 69%plusoptiX Criteria 2: Sensitivity 98% Specificity 88% |
Vision Screener Device Model Comparison
All Plusoptix Binocular Autorefractors share the same measurement technology and algorithm. Hence, measurement accuracy is identical across all available device models. The devices differ in mobility and documentation options as specified below.
Stationary in one exam room | Mobile between multiple exam rooms or locations | ||
---|---|---|---|
Vision Screener plusoptiX S16 |
Vision Screener plusoptiX S12C |
Vision Screener plusoptiX S12R |
|
Letter-size report& self-adhesive label | yes / yes | yes / yes | yes / yes |
LAN / WIFI interface | yes / yes | no / yes | no / no |
Connection to EMR / plusoptiXconnect | yes / yes | yes / yes | no / no |
External monitor interface | yes (DVI) | no | no |
Power supply | Medical power adapter (100 – 240 V AC / 50 – 60 Hz) |
6x rechargeable batteries (Standard AA size) |
6x rechargeable batteries (Standard AA size) |
Hardware Features | |||
Touchscreen operation | 4.3 Inch (resistive) | 5.7 Inch (capacitive) | 4.3 Inch (resistive) |
Weight | 26.5 oz (0.75 kg) | 35.3 oz (1.0 kg) | 28.2 oz (0.8 kg) |
Interfaces | 4 x USB, IR, DVI, LAN (RJ-45), WLAN | 2 x USB, IR, SD, WLAN | 2 x USB, IR, SD |
Identical Features | |||
Measurement Technology | Binocular infrared photo retinoscopy with unique 54 LED illumination | ||
Measurement Range | -7.00 to +5.00 dpt in 0.25 dpt increments | ||
Pupil Size | 3.0 to 8.0 mm in 0.1 mm increments | ||
Acquisition time | dynamic, in average 0.5 sec. | ||
Measuring distance | 3.3 ft ± 2 in (1 m ± 5 cm) | ||
Fixation target | Warble sound and Smiley face | ||
Certification | |||
Service Features | |||
Warranty | Hassle-free warranty (can be extended for a period of up to 10 years after initial commissioning) | ||
Operating Cost | None, because no calibration or maintenance is needed |